Selection GuideChromatography & sample prep16 min read

Pipette Tips: Universal, Filter or Low-Retention?

In one manufacturer study of fifteen tip suppliers, thirteen broke the permissible error at least once, and every one of the failing tips still fitted the pipette.

Updated November 6, 2026How Equipvia researches
Racks of disposable pipette tips in laboratory tip boxes, photographed from above with the tips standing in rows
A tip is part of the measuring instrument, not a disposable accessory: the factory calibration belongs to the pipette and tip together. Photograph via Wikimedia Commons (CC BY-SA 4.0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

Most labs buy pipette tips the way they buy paper towels: by box count, by the cheapest compatible line, and with a vague sense that they fit. That works until a 1 µL transfer starts drifting or a PCR blank comes back positive. Then the tip turns out to have been part of the measuring instrument all along.

This guide separates three decisions that usually get blurred together. First, whether a universal tip is acceptable on your pipette. Second, whether you need a filter. Third, whether low-retention tips justify their price for the liquids you handle. The evidence for each is uneven, and a good share of it comes from companies that sell tips, so the source is flagged where that matters.

The tip is part of the calibration

ISO 8655-2:2022 covers piston pipettes and sets the maximum permissible errors. The detail that matters for purchasing is that those limits apply to the pipette together with its tip. Sartorius, in an application note on tip systems, reads the standard this way: if you use tips the pipette supplier did not provide, the supplier's declaration and conformity certificate no longer apply to your setup. The same note says the standard recommends tips from the pipette's own supplier. Eppendorf's note makes the matching point that a user who wants another maker's tips should run a conformity test first.

In practice, the pipette is adjusted at the factory with the maker's tip. Change the tip and you have changed the instrument. A universal tip can still be fine. The point is that nobody has promised it is.

The gravimetric method behind these checks is laid out in ISO 8655-6:2022: dispense water onto a balance, convert mass to volume with a temperature-dependent factor, and compute systematic and random error. Nothing about it is exotic. That is why the in-house check later in this article is realistic for most labs.

What the comparison studies actually found

There are only a handful of public comparisons, and the larger ones are by tip or pipette manufacturers. Read them as evidence, not as neutral verdicts.

Eppendorf (Application Note 354). The authors tested tips from 15 manufacturers, Eppendorf included, on pipettes at 1, 10, 100 and 1,000 µL. In the tip-design analysis, 8 of the 15 suppliers exceeded the systematic error limits, and three of those also went past the wider ISO 8655 maximum. Across all four volumes, 13 suppliers broke the permissible error at least once. For 1,000 µL tips, tip length tracked the calibration result closely (R² = 0.90 once one outlier maker was left out), and some competitor tips that were longer than Eppendorf's at the same inner diameter failed. Orifice defects such as moulding flash and core shift made liquid cling or drops deflect. Their blunt conclusion was that a tip sitting securely on the cone says nothing about accuracy. Eppendorf's own tips came out best, which you would expect from a paper written by Eppendorf, and the note gives no error magnitudes for most of the failures.

Sartorius (application note SUL1006). The same pattern in a smaller test: six tip makers on 10 µL pipettes and ten on 1,000 µL pipettes. One tip was dropped after a compatibility check because it fit the 10 µL pipette badly. The note also observes that on some universal tips the sealing surface is just a thin ring around the cone, and that passing the mechanical compatibility check does not guarantee accuracy.

Thermo Fisher. Its Finnpipette note tested five sequences of ten dispenses at 1 and 10 µL and found the biggest differences at 1 µL, where the pipette with its own tip beat every universal tip. Again, all the tested tips fit.

BRAND (the counterexample). BRAND had an ISO 17025 calibration laboratory test its Transferpette S with third-party universal tips. The pipette had been calibrated with BRAND tips beforehand. BRAND's own tips gave the best accuracy, but the CV for every third-party tip it evaluated stayed inside the maker's tolerance, and it describes most of them as acceptable on higher-volume pipettes. The company's wording is deliberately cautious: "high-quality" universal tips, "particularly at higher nominal volumes."

So the sources partly disagree, and the disagreement has a pattern. Nobody finds that fit predicts accuracy. Disagreement is over how many universal tips pass, and the answer looks like it depends on the pipette volume, the tip brand and, probably, the lot. The consistent weak spot is the low end: 1 µL and 10 µL.

A check you can run yourself

If you intend to put a universal tip on a pipette that matters, test the combination. A procedure patterned on the studies above, and not a substitute for ISO 8655-6:

  1. Use a balance with suitable readability for the volume. A 1 µL dispense weighs about a milligram, so a standard analytical balance is marginal and a microbalance is better.
  2. Dispense grade 3 water at the nominal volume and at 10% of it. Sartorius ran ten dispenses per volume, and Eppendorf compared calibration with a fresh tip for every weighing against one tip reused for each volume, since the two approaches can give different answers.
  3. Convert mass to volume with the temperature-dependent factor and compare the mean to the selected volume (systematic error) and the spread (random error).
  4. Compare to the maker's specification for that pipette. If you use the tip in regulated work, record the tip brand and lot with the result.

Evaporation and drafts will spoil a 1 µL test long before anything else does. Use an evaporation trap and a closed balance chamber.

If the numbers pass, you have documented evidence the combination works, which is better than either a vendor's promise or an assumption. If they fail by a little, some pipettes can be re-adjusted for the tip you use, though that is a calibration-service question, not a bench fix.

Filter tips: required in some work, optional in the rest

A filter tip protects the pipette, not the sample. The risk is aerosol: mixing or aspirating creates tiny droplets that rise into the cone and shaft, where the next sample can pick them up.

How real is that? Eppendorf's Application Note 490 quantified it on a mechanical pipette with unfiltered tips. After one mixing step, 4 of 10 replicates had detectable genomic DNA in the pipette's air space. After 20 mixes, all 10 did. With a DNA amplicon, a single mix was enough, and the aerosol carried on the order of 10⁴ copies. In the same work, with unfiltered tips on an electronic pipette, 3 of 7 replicates showed living Pseudomonas aeruginosa inside the lower pipette and 6 of 9 showed living mycoplasma in the cone. Filtered tips blocked every sample type tested. The same note found mycoplasma persisting in roughly a third of pipette cones six weeks later, even after wiping or soaking in 70% isopropanol. Only autoclaving the lower parts cleared it.

A pipette run without filter tips also became contaminated in a gel-based PCR readout for human genomic DNA reported by one major tip manufacturer.

Two cautions on the other side.

First, not all filters are equal. A 2005 study by Rohrer and Pohlmann, from Eppendorf, built a method to count aerosol particles passing through filter tips and reported clear differences between brands. A distributor article summarising a 1999 Journal of Applied Microbiology study says that in 2,620 tests, 20% of samples showed carryover on the pipettor nose with no filter and 14% with a polyethylene filter tip. That paper itself was not reviewed for this article, so treat the figures as relayed. The same piece quotes a manufacturer statement that "barrier" is a misnomer for many tips, since only certain ones seal. Other brands advertise a self-sealing barrier and report full protection in their own carryover tests, which is a vendor claim.

Second, nobody needs filters for everything. The pattern in the sources is that filters matter for amplification-based work, nucleic acids, viral and cell culture material, and radioisotopes. For plain buffers and routine dilutions you are paying for something the sample does not need.

A practical rule: if a false positive would cost you a run, filter tips go on every pipette that touches that workflow. If not, use standard tips and keep a filter-tip box for the occasional high-risk transfer.

Low-retention tips: for particular liquids only

Water beads up on polypropylene and runs out. Glycerol creeps. So does anything with detergent in it: one major manufacturer's white paper says surfactant solutions are harder to dispense completely because they spread over more of the wall, and it groups protein and large nucleic-acid solutions with the viscous liquids. What stays on the wall is volume you aspirated but never delivered.

The vendor data is fairly consistent on direction. One white paper tested glycerol from 5% to 50% and surfactant solutions; for 0.1% Triton X-100, its low-retention tips left 0.04% behind. Another maker's data, relayed in a 2016 trade article, reported that low-retention tips held back on average 32% less liquid than the non-low-retention version for 10 µL of water and about 50% less for glycerol and BSA, with under 0.05 µL retained for 10 µL of each solution. The water number looks big in relative terms and tiny in absolute terms.

The sources also push against overuse. One manufacturer's own FAQ says that for many standard applications its non-low-retention tips may be more suitable, and recommends a pilot comparison before switching. Eppendorf's tip study found that one maker's "low retention" version performed surprisingly poorly, so the label alone guarantees nothing.

Check the additive, too. At least one major supplier states that its standard low-retention tips use a PFAS-containing additive and sells a separate PFAS-free hydrophobic line. It advises against autoclaving any of its low-retention tips.

Reach for them with 50% glycerol enzyme stocks, Tween- or Triton-containing mixes, and 1–2 µL of a precious sample. Leave them on the shelf for PBS and culture medium.

Sterile, racked or bulk

Catalogues show the split. One 250 µL low-retention line comes as 1,000 loose in a bag, while a 20 µL line in the same catalogue comes as pre-sterilised racks. Racks suit cell culture because nobody handles loose tips. Bags suit a chemistry bench where cost per tip matters more. Whichever you buy, write the lot number in the pipette log. A lot change is the first suspect when results shift for no other reason.

Putting the three decisions together

Situation Tip choice Reason from the sources
Regulated or calibrated pipette, volumes 1–20 µL Pipette maker's own tip, or a universal tip you have tested Fit does not predict accuracy; differences are largest at 1 µL
100–1,000 µL routine work Good universal tips are often acceptable; test one lot One maker found third-party CVs within tolerance; another found length-related failures at 1,000 µL
PCR, qPCR, cell or virus work Filter tips, preferably a sealing barrier type Aerosol carryover measured in two manufacturer studies
Glycerol, detergents, viscous master mix Low-retention tips Vendor data show lower residual liquid
Buffers, media, routine dilutions Standard tips Low-retention gives little benefit; filters add cost
Automated liquid handler The platform's validated rack and tip Pickup mechanisms are built around specific geometry

The last row is general practice rather than something measured in the studies above; confirm against your platform's documentation.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1600697704484 during review. CPS resolution does not verify the seller's technical claims, the tip's fit or accuracy, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

This listing bundles the three decisions into one product description: universal fit, sterile, low retention and nuclease-free, at 10 µL. Those are separate properties, and the one that matters most at 10 µL is the one not stated, which is tip length and orifice geometry. Ask for the length, the inner diameter at the orifice and any calibration data on your pipette model, and confirm what "sterile" refers to, since sterilising the rack and sterilising the tips are not the same claim.

Questions that come up

Does a snug fit mean the tip is accurate? In one 1 µL test every tip fit, and the pipette's own tip still beat all of them.

Can a filter tip be used twice? Plan on once. Aerosol was detectable after a single mixing step in the published work, so a tip already used on one sample has already met that sample's aerosol.

Do low-retention tips change the calibration? They change how much liquid leaves the tip, which is what the calibration measures. If you switch, rerun the gravimetric check rather than assuming the previous numbers still hold.

On Monday

Take the pipette with the most doubt attached, load a tip from a box nobody has validated, and weigh ten dispenses at its smallest and largest volume. A pass goes into the SOP with the brand and lot. A fail means you have found drift that the calibration sticker never mentioned.