Test Sieves: ASTM E11 or ISO 3310-1, Three Grades, and When to Replace
A compliant 150 µm sieve may contain individual openings up to 193 µm. Two sieves with the same label are not the same instrument.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.
In 1950, two researchers at the National Bureau of Standards, Frank Carpenter and Victor Deitz, ran the same material through different sets of standard sieves and found that the sets disagreed by as much as about 5 % of the total sample. Once each sieve was calibrated with glass spheres and the results were corrected to nominal openings, the disagreement fell to a few tenths of a percent. Seventy-odd years later the standards are tighter, but the underlying problem is the same: two sieves with the same label are not the same instrument.
That is what makes a test sieve an odd purchase. It looks like a commodity, a wire mesh in a brass or stainless frame, and it is bought again and again as sets wear out. Yet the label hides three layers of decision: which standard it follows, which verification grade it carries, and how its size is named. This guide takes those in turn. The firm sources are ASTM E11 and ISO 3310-1 (read only in part, as described at the end), the ASTM E2427 performance method, the ASTM B214 precision data, and the 1950 NBS paper. Much of the practical advice about wear and replacement comes from sieve makers and dealers, who sell the products involved, so it is attributed by name.
What the standards actually define
ASTM E11 covers the woven wire cloth used in test sieves and the construction of the sieves themselves. Its current edition, E11-24, applies to sieve cloth with nominal openings from 125 mm down to 20 µm. ISO 3310-1:2016 describes the same range, 125 mm down to 20 µm, and refers to ISO 565 for the nominal sizes. So for ordinary lab work the two cover the same ground, and neither covers perforated plate or electroformed sieves, which fall under other parts of ISO 3310 and other ASTM standards.
The tolerances in E11 are not a single number. The abstract says it specifies three: the allowed variation of the average opening, the maximum variation of any single opening, and the maximum standard deviation of the openings. Working from the E11-15 table for an example, with the caveat that the 2024 edition may differ, a nominal 150 µm opening (the "No. 100" sieve) allows the average to vary by ±6.6 µm, no single opening may exceed 43 µm over nominal, giving a largest permitted individual opening of 193 µm, and the typical wire diameter is 0.1 mm with a permitted range of 0.085 to 0.115 mm. In other words, a compliant 150 µm sieve can contain openings nearly a third larger than its name. The practical meaning of "150 µm" is a population of openings centred close to 150 µm, not a guarantee about each particle.
Two pieces of arithmetic follow, both arithmetic and not from any standard. First, with 150 µm openings and 100 µm wire, the open area is (150 / 250)², about 36 %, so fine sieves have little open area to spare and sample load matters. Second, the E11-15 table shows the tolerance getting relatively looser as the sieve gets finer. The permitted average variation is ±8.7 µm at 212 µm, ±7.6 at 180, ±6.6 at 150, ±5.8 at 125 and ±5.2 at 106, which works out to 4.1 %, 4.2 %, 4.4 %, 4.6 % and 4.9 % of nominal. The largest permitted single opening runs from about 1.25 times nominal at 212 µm to about 1.33 times at 106 µm. If your specification sits near a boundary on a fine sieve, expect more sieve-to-sieve scatter than the same boundary on a coarse one.
The older paper makes a related point from the opposite direction. Carpenter and Deitz report that effective openings, as measured with glass spheres, ran larger than nominal; their example, a sieve with a 350 µm nominal opening, was about 6.3 % larger. They also note that the average-opening tolerance of their day was ±3 % for No. 16 and coarser sieves, rising to ±7 % for No. 200 and finer, and that in a √2 series two neighbouring sieves could, in the worst case, differ by roughly 5 % to 33 % in the size they actually separate. Those tolerances are not today's, but the pattern, in which fine sieves vary more in relative terms, matches the E11-15 numbers above.
The three grades: compliance, inspection, calibration
E11 sells the same cloth at three levels of proof. A compliance sieve is made from cloth that was inspected before it was mounted, by sampling the roll. Inspection and calibration sieves are measured after mounting, individually, and the standard says calibration sieves have at least twice as many openings measured as inspection sieves. The statistical confidence behind the standard deviation check rises with the grade: roughly 66 % for compliance, 99 % for inspection and 99.73 % for calibration, per the E11-15 text. Gilson, a sieve dealer, describes the same ladder, with inspection sieves checked on an optical comparator against NIST-traceable measurements.
Two cautions apply. First, none of the grades tells you the effective opening that particles will see. A dealer article says so itself: the mean opening of a calibrated sieve can lie within tolerance but still differ slightly from nominal. A trade article in Painting & Coating Industry makes a blunter point, that the inspected openings are less than 0.001 % of the openings on a sieve, and promotes an "effective aperture" measured with glass microspheres instead. That is a vendor-flavoured claim, but it echoes what Carpenter and Deitz proposed in 1950.
Second, a higher grade is not automatically the right purchase. Compliance sieves are reasonable for working sets where results are internal and compared against a reference set, and inspection or calibration grade is worth the premium for the one reference set, a "master stack" in the language of sieve makers, that you keep out of daily use and use to judge the others. W.S. Tyler and Haver & Boecker both recommend this master-stack approach in their technical blogs.
Gilson also says E11 and ISO 3310-1 have "similar requirements" and both use compliance, inspection and calibration classes. That could not be confirmed from the ISO text available, so treat it as a dealer's summary. What can be confirmed is the E11 side.
Naming: mesh, number and micron
Sieve names are where reorders go wrong. E11 uses a standard designation in millimetres or micrometres and an "alternative" designation, the familiar No. 100, which the standard itself says is only approximate. A 150 µm opening is "No. 100", 125 µm is "No. 120", and 106 µm is "No. 140". Tyler's older mesh scale numbers differ for some sizes: Retsch's comparison table, for example, lists 106 µm as Tyler 150 while the US number is 140. A wire cloth dealer's table lists 150 µm as "100 mesh" and 125 µm as "115 mesh". Hence "100 mesh" alone could mean different openings depending on who is selling.
The ISO series adds sizes that the US series does not have. Gilson's comparison lists ISO 565 sizes such as 160, 140, 112 and 100 µm, which sit between the E11 sizes of 180, 150, 125 and 106 µm. So an ISO 140 µm sieve is not a substitute for a 150 µm E11 sieve, and a method written around a US number should be bought by its micron value.
Frame size is another point where the standards agree and the catalogues blur. E11-15 lists the 8 inch frame as 203 mm, the size ASTM B214 specifies for its sieves, and the ISO frame diameter could not be found in the sources available, so check the frame against your shaker before ordering.
Verifying and replacing: what the evidence supports
Wear is the part of the purchase that never appears on the invoice. Gilson notes that wire cloth stretches, sags and tears with age, and Haver & Boecker says two factors mainly compromise agreement between old and new sieves. Carpenter and Deitz found that wire diameter variation within specification mattered little, while uneven openings, such as those in a worn or distorted sieve, were the larger problem.
There is no way to adjust a sieve, so verification means comparison. ASTM E2427-22 is designed for this: it compares a working sieve with an inspection or calibration reference sieve using a reference material, usually calibrated glass beads, and the standard's own scope says it is a way to check accuracy and long-term reliability because the measuring capability of a sieve cannot be adjusted. Gilson adds that the beads are expensive and can be used only a limited number of times, and that a sieve whose performance drifts can be sent for optical examination or simply replaced.
The best evidence that sieve-to-sieve differences matter in practice is ASTM B214's precision section. In its interlaboratory study, twelve laboratories each made three tests on four metal powders using two sieve sets: one set that was circulated among them, and one each laboratory chose from its own stock. The between-laboratory reproducibility limit was smaller for the circulated sieves than for the in-house ones. That is a pointed result for anyone who buys sieves one at a time: shared sieves agreed better than sieves each laboratory chose from its own stock.
The reading here is that a calendar interval for replacement is hard to justify from these sources. None of them gives one. What they support is a trend: run the same reference material on the working sieve now and then, plot the result, and retire a sieve when it drifts or when it is damaged.
Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 62431039404 during review. CPS resolution does not verify the seller's technical claims, the grade, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.
An accuracy figure of ±5 % is not a grade. E11 grades are statements about how many openings were measured and to what confidence, not a single percentage on the frame, and a ±5 % figure is wider than the permitted average variation at any of the fine sizes in the table above. Brass also wears faster than stainless in abrasive service, and soft frames distort when a stack is over-tightened. For soil and aggregate screening that may be entirely adequate, but a set like this should not become the master stack that other sieves are judged against, and the micron value of each opening still needs to be confirmed in writing.
A short reorder routine
This is a condensed version of the above, not a published procedure.
- Specify the opening in micrometres or millimetres and the standard (E11 or ISO 3310-1), and treat the mesh number as a convenience.
- Choose the grade by role: a reference set at inspection or calibration grade, working sets at compliance grade if you will monitor them.
- Match the frame diameter and height to the shaker before ordering, since stacks must nest.
- Brush and inspect after each use. B214 asks for back-brushing to dislodge wedged particles and for a mass balance of at least 99 % of the test portion, which is a cheap warning sign for a damaged screen or pan.
- Keep a dated record of performance runs for every sieve in a stack.
What else to check
The sieve is only half the method. ASTM B214 fixes sieving time at 15 minutes and lets friable materials use less if the time is reported. Carpenter and Deitz proposed instead finding an end point by the sieving rate, about 0.1 g per minute, and starting with a trial near 10 minutes for 100 g. A sieve-maker blog advises that you cannot sieve too long but can sieve too briefly. These are different philosophies, and the right one depends on whether your method states a time.
Sample size is the other half of that conversation. Carpenter and Deitz preferred 100 to 150 g of material and said 50 to 500 g could work, with very small samples giving erratic results and very large ones blinding the sieve. They also found that 1 to 2 % moisture had a negligible effect, though they still advised drying the sample first. That was bone char sieved on a tapping shaker, so the numbers do not transfer to a fine pharmaceutical powder on an air-jet sieve. The principle does: if your results wander between runs on the same sieve, check load and time before blaming the mesh.
What this article could not check
ISO 3310-1 could not be read beyond its scope and contents, so statements about its tolerances and classes are limited to what dealers and summaries say. The E11 values come from the 2015 edition's text and tables, not the current 2024 edition, which was read only as an abstract. The E11 formulas were garbled in the copy used, so only table values are quoted. ASTM B214 was read from a sample document, with the repeatability equation illegible. Sieve-maker and dealer pages come from firms with an interest in selling sieves and reference materials.