Calibration Weights: OIML Class, Tolerance and Recalibration

At one kilogram, E2 allows ±1.6 mg and F2 allows ±16 mg. Each class step is about a factor of three, which is what you actually pay for.

Updated November 8, 2026How Equipvia researches
A set of metal calibration weights photographed together with their individual identification markings visible
The class sets the tolerance and the certificate says where inside it the weight actually sits. Modern classes also limit density and magnetism. Photograph via Wikimedia Commons (CC BY 4.0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

Look at the 1 kg row of the OIML table. E2 allows ±1.6 mg. F2 allows ±16 mg. Two weights, the same stainless steel, near-identical cases, and the certificate on one promises ten times less error than the other. Lab buyers often choose between them by catalogue order. This piece tries to replace that habit with something firmer: what the classes mean, what USP asks of a test weight (less than most people think), where published tables contradict each other, and how long a set stays good.

A note on sources. The class tables and pharmacopoeial text are standards. Advice on handling, intervals and matching class to balance mostly comes from weight sellers, and they are named. Anything that is opinion is labelled as such.

What the classes are

The nine classes in R 111-1 (E1, E2, F1, F2, M1, M1-2, M2, M2-3, M3) run from 1 mg to 5,000 kg, with E1 the tightest. For laboratory standards you will mostly meet E1 through F2, a point an OIML guidance document (E 3) makes directly. Whatever the class, the calibration lab's own uncertainty has to stay at or below a third of the weight's maximum permissible error (MPE), and a Troemner handbook quotes the same ratio for ASTM E617. E1 weights ship with a certificate every time, says Rice Lake.

The US family is ASTM E617, classes 0 to 7 with finer 000 and 00 above them. At 100 g, Troemner lists Class 1 at 0.25 mg and Class 2 at 0.50 mg. Set those beside OIML E2 (0.16 mg) and F1 (0.5 mg) and the pairing is easy: E2 is a little tighter than ASTM 1, and F1 lands on ASTM 2. Rice Lake's cross-reference chart puts E2 alongside ASTM 0 and 1 and F1 alongside 2, which agrees.

Now the 2004-edition limits, as Gram Group publishes them (VWR's catalogue matches for F1). At 100 g: ±0.05 mg for E1, ±0.16 mg for E2, ±0.5 mg for F1, ±1.6 mg for F2 and ±5 mg for M1. At 200 g the figures are ±0.10, ±0.3, ±1.0, ±3.0 and ±10 mg. At 1 kg they are ±0.5, ±1.6, ±5, ±16 and ±50 mg. Reading across, each step from E2 to F1 to F2 to M1 is about a factor of three in allowed error, which is an observation from these rows. Skipping one class therefore costs roughly three times the tolerance, and skipping two about ten.

Where the tables disagree

Printed tolerance charts are not all the same. Genesee Scientific's ASTM-and-OIML sheet gives E2 at 1 kg as ±1.5 mg and F2 as ±15 mg, and so does a Scribd copy of an OIML-style chart. Gram Group, VWR's weight listings and The Engineer say ±1.6 and ±16. The second set is what R 111-1:2004 appears to print, with the rounder numbers coming from an older or simplified table, but the standard could not be opened to check. A 6 % gap is small beside the threefold step between classes; it still matters for a certificate that lands near its limit. OIML itself amended R 111-1 in 2025 to correct a table in Annex B, so lift tolerances from the current standard or the certificate, not from a PDF that was forwarded to you.

Vendor pages slip as well. Troemner's classification page works an example with ASTM Class 1 at 100 g equal to 0.025 mg, then lists the same class at 0.25 mg a few lines later. Genesee's table backs the larger figure. A misplaced decimal like that moves a weight two classes, so do not copy a tolerance from a product page without a second source.

Choosing a class for the balance

USP <41> sets the accuracy test at 0.10 % of the test weight. The weight must fall between 5 % and 100 % of the balance capacity, and its own MPE or calibration uncertainty may be no more than a third of that limit. A 200 g balance checked with a 100 g weight gives a limit of 100 mg and a weight allowance of about 33 mg, by arithmetic. Against that, a 100 g weight is ±5 mg in M1, ±1.6 mg in F2 and ±0.16 mg in E2. All three clear the bar. One industry publication concludes that class F is usually enough for this test and E2 is not always mandatory.

Troemner would still have you buy tighter than that. A weight that drifts out of tolerance between calibrations puts every result since the last good check under investigation, and a finer class leaves room for the drift. It is a seller's argument, and a sound one.

For a general rule on matching class to balance, Troemner's selection guide lists, for an analytical balance reading 0.1 mg, classes from ASTM 1 through OIML E2 and F1; for precision balances reading 0.05 g to 0.001 g, F1 and F2; and for medium-resolution balances reading 1 to 2 g, M1 and M2. A working rule, which is not published anywhere, is to keep the test weight's tolerance within a handful of readout steps. For a 0.1 mg analytical balance, a 200 g F1 weight (±1 mg, ten steps) is too loose for strict work, while E2 at ±0.3 mg (three steps) is a fair match. If your SOP allows looser limits, F1 is fine.

Numbers make this concrete. On a 220 g balance reading 0.1 mg, a 200 g E2 weight at ±0.3 mg is three readout steps, a 100 g E2 at ±0.16 mg is under two, and a 50 g E2 at ±0.10 mg is one. The same weights in F1 are ten, five and three steps. A selection chart in German that circulates as an OIML R 111 explainer takes a similar line from the other direction: it maps capacity-to-readability ratio to class, with analytical balances above 100,000 divisions at E2, ratios up to 100,000 at F1, up to 50,000 at F2 and up to 10,000 at M1. That chart could not be traced to its publisher, so treat it as a heuristic. By its logic, 220 g over 0.1 mg is 2.2 million divisions, comfortably E2 territory.

Reading the certificate

A weight's class is a promise about its limits. The certificate tells you where it actually sits inside them. Troemner's USP <41> page says test weights need a traceable calibration report listing the correction and the measurement uncertainty, and NIST's double-substitution procedure expresses results as conventional mass, referenced to an air density of 1.2 kg/m³. So for each weight the certificate gives a correction and an expanded uncertainty, and the second should be a fraction of the MPE.

Here is where it pays to be strict, as a decision rule rather than a published one. Add the correction's size and the uncertainty together. If the sum still sits well inside the class tolerance, the weight is doing its job. If it already uses up most of the tolerance on the day it arrives, there is no room for drift, and the first recalibration is likely to fail it.

What the weight is made of

Weights are not only a tolerance. OIML sets limits on density, magnetism and construction for each class, and they matter on a high-resolution balance.

Rice Lake lists density for its E1 and E2 stainless weights as 7.95 g/cm³, and 7.84 g/cm³ for F1, F2 and M1, with aluminium at 2.7 g/cm³ for the 1 to 5 mg pieces. Density matters because the balance weighs in air. The Troemner handbook states the OIML requirement as a density such that a 10 % change from the reference air density of 1.2 kg/m³ produces no more than a quarter of the MPE in error. In ordinary rooms that is comfortably met.

Magnetism limits tighten with class, and susceptibility limits are listed separately by mass. The handbook's table puts the maximum magnetisation at 3 µT for E1, 10 µT for E2, 30 µT for F1 and 100 µT for F2.

Construction differs as well. According to the same handbook, E1 and E2 weights are one-piece with no adjusting cavity, while one distributor's description of its F1 weights mentions an adjusting cavity except in the 1 to 20 g monobloc pieces.

Sets or single weights

A full set is rarely what routine checks need. The USP test uses one weight inside the 5 % to 100 % window, and repeatability does not require a small one. Rice Lake's 100 g range sets show how quickly the piece count grows: a 100 g to 1 mg set has 21 pieces for 211.11 g, a 100 g to 100 mg set has 13 pieces, and a 100 g to 1 g set has 9. A practical suggestion is to buy the single weights your SOP actually names, such as one near capacity and one near the lower end of the working range, and carry a full set only if you need to check linearity across the whole range. Fewer pieces also means fewer things to handle and to send out for calibration.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 60568483263 during review. CPS resolution does not verify the seller's technical claims, the class, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

An F1 label on a listing is a claim about the tolerances above, and the purchase is only complete with the paperwork. Ask for the calibration certificate, whether the stated uncertainty is below a third of the MPE, and whether the certificate names the individual serial number. A weight sold without a traceable certificate is a reference mass of unknown status, whatever the title says.

Handling and storage

Handling is where weights are most often damaged by their owners. Mettler Toledo's weight-handling paper says fingerprints and the residue they leave can add up to 50 µg, and recommends non-magnetic, non-abrasive tweezers or synthetic or leather gloves, adding that clean cotton gloves are acceptable for looser tolerances. NIST's GMP 10 says small weights should never be handled by hand, and notes that forceps and gloves themselves must be cleaned. One trade guide makes the exception that only cast-iron weights are meant to be handled bare.

Mettler Toledo's calibration-service brochure adds that cleaned weights need a stabilisation period before calibration so the surface is stable, which is worth remembering if you clean a weight yourself and test it straight away. Two additions, not from those sources: dedicate one pair of forceps to the reference set and keep it in the case, and let weights acclimatise to the balance room before use.

How often to recalibrate

Here the sources do not agree, and it is better to show the spread than pick a winner.

Source Guidance
NIST No set interval; base it on measurement data. For a 100 g weight set, typical intervals run from 6 months to 5 years or longer
Troemner Once a year for infrequent to moderate use; interval set by observed drift
Scales Galore No industry standard; annually is recommended
Mettler Toledo (a brochure chart) 1 to 2 years for customer working weights

NIST's own pages are the most careful. They say the right interval depends on accuracy requirements and on how the weight actually behaves, and that each calibration report should include as-found data, so the next interval can be shortened or lengthened on evidence. The 2025 revision of USP <41> pushes in the same direction, according to a Mettler Toledo white paper: calibration must include measurement uncertainty and as-found and as-left results.

A routine offered as a suggestion: start at one year, keep every certificate, and compare each as-found value with the previous as-left value. If the change is under a quarter of the MPE after two cycles, lengthen the interval; if it is a half or more, shorten it and look at how the weight is handled. Between sends, compare the working set against a reference-class weight kept only for that purpose, every few months. Plan around the downtime as well: a calibration lab may keep a set for days, and the balance still needs its routine checks in that gap, which is one more reason to hold a small second set.

When to suspect the weight

A balance that passes with one weight and fails with another of the same nominal value points at the weight. So does a weight that reads consistently high by about the same amount, which fits fingerprint residue or dirt, and one that has changed since the last certificate by more than its tolerance. Visible pitting, scratches or a dull surface are reasons to send it out regardless of date. A clean way to separate balance from weight is to compare against a second weight of the same nominal value from a different set. If the two disagree by more than their combined tolerances, one of them is out.