Selection GuideWorkshop & machinery15 min read

Volumetric Pipettes: Class A, AS or B, and the ASTM E969 against ISO 648 Difference

At 1 and 5 mL the ISO figures are 1.33 and 1.50 times the ASTM ones. From 10 mL upward the two standards print the same numbers.

Updated November 15, 2026How Equipvia researches
Several glass volumetric pipettes of different sizes laid out together on a bench
A transfer pipette delivers one volume, and the class letter plus the standard behind it is what sets the tolerance. Photograph via Wikimedia Commons (CC BY-SA 4.0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

A 1 mL Class A volumetric pipette is ±0.006 mL under ASTM E969 as sold by Corning and Fisher, and ±0.008 mL under ISO 648 as sold in the MBL Class AS range. A 5 mL pipette is ±0.010 mL against ±0.015 mL. At 10, 25, 50 and 100 mL the two sets are the same: ±0.02, ±0.03, ±0.05 and ±0.08 mL. The ISO figures are therefore 1.33 and 1.50 times the ASTM figures at 1 and 5 mL and equal from 10 mL up. ISO 648:2008 adds a Class AS with a 5 s waiting time, and ASTM E969 has two classes, A and B.

Tolerances by volume

ASTM E969-02(2012) covers two classes, Class A (precision) and Class B (general purpose). The ASTM column below is the Table 1 text from the standard's abstract page, cross-checked against Corning, Kimble and Sigma-Aldrich Class A listings. The ISO columns are the 1977 edition of ISO 648, whose full table was available as a sample. ISO 648:2008 is the current edition, and its summary gives two examples, 10 mL Class A ±0.02 mL and 1 mL Class A ±0.008 mL, which match the 1977 table.

Nominal, mL ASTM E969 Class A ASTM E969 Class B ISO 648:1977 Class A ISO 648:1977 Class B
0.5 ±0.006 ±0.012 ±0.005 ±0.010
1 ±0.006 ±0.012 ±0.008 ±0.015
2 ±0.006 ±0.012 ±0.010 ±0.020
5 ±0.010 ±0.020 ±0.015 ±0.030
10 ±0.020 not in excerpt ±0.020 ±0.040
20 ±0.030 ±0.060 ±0.030 ±0.060
25 ±0.030 ±0.060 ±0.030 ±0.060
50 ±0.050 ±0.100 ±0.050 not legible
100 ±0.080 ±0.160 ±0.080 ±0.150

The ISO 1977 table also lists 200 mL, but the printed values were not legible in the sample. Fisherbrand lists a 200 mL serialised Class A pipette at ±0.16 mL.

Across the nine volumes in the table, ISO Class A equals ASTM Class A at five volumes (10, 20, 25, 50 and 100 mL), is looser at three (1, 2 and 5 mL) and is tighter at one (0.5 mL). The ratio of ISO to ASTM Class A is 0.83 at 0.5 mL, 1.33 at 1 mL, 1.67 at 2 mL, 1.50 at 5 mL and 1.00 from 10 mL up. The ratio of Class B to Class A is 2.00 in ASTM at every volume in the excerpt. In ISO 1977 it is 2.00 at most sizes, 1.88 at 1 mL and 1.88 at 100 mL.

Class AS and the vendor listings

Marienfeld states that Class AS tolerances are half of the Class B tolerances under DIN ISO. At 5 mL that is half of ±0.030 mL, which is ±0.015 mL, the figure Marienfeld, MBL and one other supplier print for 5 mL Class AS. The MBL range lists 1 mL at ±0.008, 2 mL at ±0.010, 5 mL at ±0.015, 10 mL at ±0.020, 15, 20 and 25 mL at ±0.030, 50 mL at ±0.050 and 100 mL at ±0.080. Two listings disagree with that set: one prints ±0.007 mL for 1 mL against ±0.008 mL at Marienfeld and MBL, and Marienfeld's certified table prints ±0,0010 ml for 2 mL where its uncertified table prints ±0,010 ml, which looks like a typo.

Corning and Kimble list ASTM Class A borosilicate pipettes at ±0.006 mL for 1 and 2 mL, ±0.01 mL for 3, 4 and 5 mL, ±0.02 mL for 10 mL, ±0.03 mL for 15, 20 and 25 mL, ±0.05 mL for 50 mL and ±0.08 mL for 100 mL. Kimble also lists 0.5 mL at ±0.006 mL and states that its 75 and 200 mL sizes are designed from ASTM E542, the calibration practice, and not from E969. Sibata sells wide-tip pipettes to the Class B tolerances of E969.

Waiting time and outflow time

ISO 648:1977 describes two Class A options: no definite waiting time, in which about 3 s is observed, or a 15 s waiting time. Class B should have no definite waiting time. ISO 648:2008 introduces Class AS with a 5 s waiting time that must be marked on the pipette, and its list of changes says the Class AS waiting time was reduced to 5 s. The summary of the 2008 edition lists the required markings as nominal volume, reference temperature, "Ex", accuracy class, waiting time where applicable, and the manufacturer's mark. The adjustment water is grade 3 under ISO 3696, and the 1977 edition defines capacity as water at 20 °C.

The excerpt of ASTM E969 says the pipette is emptied with the tip against the vessel and that no after-drainage period is allowed. The ASTM and ISO texts therefore differ: 0 s after-drainage against about 3 s or 5 s or 15 s of waiting. Neither standard was read in full, so the exact wording of each rule is not confirmed here.

Minimum outflow time in the ASTM table, Class A: 8 s at 5 mL, 25 s at 20, 25 and 50 mL, and 30 s at 100 mL. The maximum outflow time for either class is 60 s. Fisherbrand's listed minimum flow times match the Class A column: 5 s at 0.5 mL, 8 s at 1, 2 and 5 mL, 10 s at 6 to 9 mL, 25 s at 20 to 50 mL and 30 s at 100 mL. A 100 mL Class A pipette therefore has a 30 to 60 s delivery window, a range of 2.0 times.

NIST Handbook 133 requires a Class A Mohr or Class A transfer pipette for high-precision volumetric measurement and tells the user to touch the tip to the wet sidewall of the receiving container so surface tension aids full drainage. NIST bases its pipette description on ASTM E969-02, reapproved 2012.

Relative error and the mass a balance must resolve

Relative error is the tolerance divided by the nominal volume. Mass equivalent uses water at 0.9982 g/mL for 20 °C.

Nominal, mL Tolerance, mL (ASTM A) Relative error Mass equivalent, mg ISO A, mL ISO relative error
1 ±0.006 0.60 % 6.0 ±0.008 0.80 %
2 ±0.006 0.30 % 6.0 ±0.010 0.50 %
5 ±0.010 0.20 % 10.0 ±0.015 0.30 %
10 ±0.020 0.20 % 20.0 ±0.020 0.20 %
25 ±0.030 0.12 % 29.9 ±0.030 0.12 %
50 ±0.050 0.10 % 49.9 ±0.050 0.10 %
100 ±0.080 0.08 % 79.9 ±0.080 0.08 %

A 25 mL delivery of water at 20 °C weighs 24.955 g. A 0.1 mg balance resolves 1/60 of the 1 mL Class A tolerance of 6.0 mg and 1/299 of the 25 mL tolerance of 29.9 mg. The 1 mL pipette at ±0.60 percent is 7.5 times the relative error of the 100 mL pipette at ±0.08 percent.

ASTM Class B at the same volumes works out at 1 mL ±0.012 mL for 1.20 percent and 12.0 mg, 5 mL ±0.020 mL for 0.40 percent and 20.0 mg, 25 mL ±0.060 mL for 0.24 percent and 59.9 mg, and 100 mL ±0.160 mL for 0.16 percent and 159.7 mg. The Class B 25 mL figure of 0.24 percent is 2.0 times the Class A 0.12 percent, and the Class B 100 mL tolerance of ±0.160 mL equals the ±0.16 mL that Fisherbrand lists for a 200 mL Class A pipette.

Combining pipettes and intermediate sizes

ASTM Class A combinations: 5 mL plus 10 mL gives 15 mL at a worst case of ±0.030 mL (0.20 percent) and a root-sum-square of 0.0224 mL (0.149 percent). One 15 mL pipette is ±0.030 mL (0.20 percent). Ten transfers of 1 mL give a worst case of ±0.060 mL (0.60 percent) and a root-sum-square of 0.019 mL (0.19 percent), against ±0.020 mL (0.20 percent) for one 10 mL pipette and ±0.030 mL (0.12 percent) for one 25 mL pipette.

Kimble lists 22 Class A sizes from 0.5 to 100 mL with six distinct tolerances: ±0.006, ±0.01, ±0.02, ±0.03, ±0.05 and ±0.08 mL. Listed intermediate sizes include 1.5 mL at ±0.006, 9 mL at ±0.02, 12 mL at ±0.02, 30 mL at ±0.03, 40 mL at ±0.05 and 75 mL at ±0.05 mL. Fisherbrand lists 7 mL at ±0.01 and 8 mL at ±0.02, so the tolerance doubles between adjacent 1 mL steps. The sizes that ISO 648 itself lists are 0.5, 1, 2, 5, 10, 20, 25, 50, 100 and 200 mL: ten sizes against 22 on the Kimble list.

Graduated pipettes are a separate standard

ISO 835-1:1981 splits graduated pipettes into three parts: Part 2 for no waiting time (Class A and B), Part 3 for a 15 s waiting time (Class A only) and Part 4 for blow-out (Class B only). NIST names the Class A Mohr (graduated) pipette and the Class A transfer pipette as the two accepted for high-precision volumetric work, and its example dispenses 3.2 mL from a graduated pipette by reading the scale. A volumetric pipette delivers one volume, while the graduated pipette adds a reading error at the scale. No tolerance values for graduated pipettes could be read cleanly, so no comparison is made here.

Colour codes

Colour bands in the MBL, midsci and Sigma-Aldrich listings: 1 mL blue, 2 mL orange, 5 mL white, 10 mL red, 20 mL yellow, 25 mL blue, 50 mL red, 100 mL yellow. MBL adds 15 mL green, and one supplier lists 3 mL black. Repeated colours are blue (1 and 25 mL), red (10 and 50 mL) and yellow (20 and 100 mL), which is three colours for six sizes. In the ASTM Table 1 excerpt the colour text was jumbled, so it was not used.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1601725634228 during review. CPS resolution does not verify the seller's technical claims, the class, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

For a pipette the missing line decides whether the class claim means anything. "Class A" alone covers ±0.006 mL at 1 mL under E969 and ±0.008 mL under ISO 648, and it says nothing about which waiting time the calibration used, since ISO 648:2008 requires that to be marked and ASTM allows no after-drainage period at all. Ask which standard the pipette is made to, the tolerance for each size you are buying, whether it carries an individual serial number and certificate, and what waiting time the certificate was based on. Without those four answers the class letter is a label rather than a specification.

Glass, graduation and listed prices

Corning lists borosilicate glass with enamel graduations in sizes 1 to 25 mL, colour-coded. MBL lists soda-lime glass with amber stain graduations fused into the glass that resist aggressive cleaning. Fisher Ireland states that MBL's Class AS range with certification carries an individual serial number and a calibration certificate traceable to national standards. Marienfeld states its stain graduation penetrates the glass and is more resistant than a fused blue enamel graduation, and notes its range follows DIN EN ISO 648 except the 3, 4 and 15 mL sizes.

Fisher lists the Corning Class A case of 12 at $329.00 for 1 mL, $325.00 for 10 mL, $490.00 for 15 mL, $559.00 for 25 mL, $929.50 for 50 mL and $1,015.00 for 100 mL. Per pipette that is $27.42 at 1 mL, $27.08 at 10 mL, $40.83 at 15 mL, $46.58 at 25 mL, $77.46 at 50 mL and $84.58 at 100 mL. The case price at 100 mL is 3.09 times the case price at 1 mL for 100 times the volume, which is the practical argument for keeping small sizes in stock and buying the large ones sparingly.

Size ranges per catalogue

Corning lists 11 Class A products: 1, 2, 3, 4, 5, 10, 15, 20, 25, 50 and 100 mL. Fisherbrand lists 22 volumes from 0.5 to 200 mL. MBL lists 9 Class AS sizes: 1, 2, 5, 10, 15, 20, 25, 50 and 100 mL. Marienfeld lists 12: 0.5, 1, 2, 3, 4, 5, 10, 15, 20, 25, 50 and 100 mL. Against the 10 sizes in the ISO 648:1977 series, MBL covers 8 (it lacks 0.5 and 200 mL) and adds 15 mL, while Marienfeld covers 9 (it lacks 200 mL) and adds 3, 4 and 15 mL, which it says are outside DIN EN ISO 648. The ASTM Table 1 lists capacities as small as 0.5 mL and as large as 100 mL in the excerpt, with 3, 4, 6, 7, 30 and 40 mL among them.

What this article could not check

The full text of ASTM E969-02 was not read beyond the abstract table, nor was ISO 648:2008 in full. The Class B value for the 10 mL pipette, the 1977 Class B value at 50 mL, and any test showing how much a 5 s or 15 s waiting time changes the delivered volume were not found. The mass equivalents above are tolerances converted to water mass, not measured delivery errors.

Share this article

View all guides