Micropipette Limits by Volume: ISO 8655-2, the 2002 against 2022 Editions, and the Gravimetric Test

The absolute limits stay the same at every volume a pipette can be set to. At 10 percent of nominal, an 8 µL systematic limit becomes 8 percent.

Updated November 16, 2026How Equipvia researches
Several single-channel micropipettes hanging in a rack on a laboratory bench
A pipette on its own has no tolerance: ISO 8655 treats the instrument and its tip as one system. Photograph via Wikimedia Commons (CC BY-SA 4.0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

ISO 8655-2:2002, nominal 1000 µL, type A. Systematic limit: ±0.8 percent, ±8.0 µL. Random limit: ±0.3 percent, ±3.0 µL. The text gives the same absolute limits at every selectable volume in the useful range. At 500 µL they are 1.6 and 0.6 percent. At 250 µL they are 3.2 and 1.2 percent. At 100 µL they are 8.0 and 3.0 percent. Gilson publishes 3 µL at 100 µL and 8 µL at 1000 µL for the PIPETMAN P1000, and its ISO summary gives 8 µL at all volumes.

ISO 8655-2:2002 limits by nominal volume

Table 1 of ISO 8655-2:2002 covers piston pipettes of type A (air displacement) and type D1 (positive displacement with a fixed piston). The values below are from the Mettler Toledo quick-check procedure, which reproduces the table, and agree with the rows visible in a scanned copy of the standard for 5, 10, 20 and 50 µL.

Nominal, µL Systematic, ±% Systematic, ±µL Random, ±% Random, ±µL
2 4.0 0.08 2.0 0.04
5 2.5 0.125 1.5 0.075
10 1.2 0.12 0.8 0.08
20 1.0 0.2 0.5 0.1
50 1.0 0.5 0.4 0.2
100 0.8 0.8 0.3 0.3
200 0.8 1.6 0.3 0.6
500 0.8 4.0 0.3 1.5
1000 0.8 8.0 0.3 3.0
2000 0.8 16 0.3 6.0
5000 0.8 40 0.3 15.0

The systematic error is the deviation of the mean of a tenfold measurement from the nominal or selected volume. The random error is the coefficient of variation of the tenfold measurement, or the repeatability standard deviation in µL. A footnote in the scanned copy says that for type D1 the limit at 100 and 200 µL may be ±0.4 percent. For intermediate nominal volumes, the values of the next greater listed volume apply. The standard's own example is a 25 µL pipette, which takes the 50 µL values: ±0.5 µL systematic and ±0.2 µL random. That is 2.0 percent and 0.8 percent of 25 µL against 1.0 percent and 0.5 percent for a 20 µL pipette.

The same limits at 10 and 50 percent of nominal

The 2002 text states that the absolute limits of the nominal volume apply to every selectable volume. For a 10 to 100 µL pipette, nominal 100 µL, the systematic limit is ±0.8 µL and the random limit ±0.3 µL at every volume, which as a percentage of the setting works out as follows.

Nominal, µL At 10 % of nominal, sys / random At 50 % of nominal, sys / random
10 12.0 % / 8.0 % 2.4 % / 1.6 %
20 10.0 % / 5.0 % 2.0 % / 1.0 %
50 10.0 % / 4.0 % 2.0 % / 0.8 %
100 8.0 % / 3.0 % 1.6 % / 0.6 %
1000 8.0 % / 3.0 % 1.6 % / 0.6 %
5000 8.0 % / 3.0 % 1.6 % / 0.6 %

From 100 µL nominal upward the table is identical in percent: 8.0 and 3.0 at 10 percent of nominal, and 1.6 and 0.6 at 50 percent. The ratio of the systematic to the random limit is 2.67 for every nominal volume from 100 µL up and 2.0 at 2 and 20 µL.

What the 2022 revision changed

The ISO 8655:2022 series replaced the 2002 series. The EURAMET report on the revision says the maximum permissible errors are now given only as percentages, the tables are organised by ranges of nominal volume, and for each nominal value the limit at 10, 50 and 100 percent is described. Eppendorf and Gilson both say this removes the need for the user to work out the 10 and 50 percent limits. Multichannel pipettes get their own table; in 2002 their limits were twice the single-channel values, so a 1000 µL multichannel pipette was ±1.6 percent and ±0.6 percent. The EURAMET report also lists an informational annex on motorised micropipettes and an extension of the covered range up to 20 mL.

The Eppendorf comparison table, in the rows that could be read: type D2 systematic limit at 20 to 100 µL, 1.5 percent in 2002 against 1.4 percent in 2022; one random limit at 10 to 20 µL, 0.8 percent against 0.5 percent; and no change to 0.8 and 0.3 percent above 100 µL. Eppendorf states that some limits were lowered. Gilson and Eppendorf both state that the pipette and its tip are one system. The 2022 limits were not rebuilt by volume here, because the table extraction came through partly scrambled, and the text of ISO 8655-2:2022 was not read.

Vendor specifications against the ISO limit

Gilson publishes tighter numbers than ISO for most volumes. For the PIPETMAN M P10M the specification page lists systematic errors of ±0.040 µL at 0.5 µL, ±0.025 µL at 1 µL, ±0.060 µL at 5 µL and ±0.080 µL at 10 µL, against ±0.120 µL at every volume for ISO. At 1 µL that is 2.5 percent against 12 percent, a ratio of 4.8. At 0.5 µL Gilson's 8.0 percent is one third of the 24 percent that ±0.12 µL represents. Gilson also notes that ISO gives limits at 10, 50 and 100 percent only, while its own table includes the minimum volume of 0.5 µL on the P10.

Those tighter figures hold only with Gilson's recommended DIAMOND tips, per the same page. Under the 2022 text the pipette and tip are one system, so a vendor figure measured with the maker's own tip does not carry over to a third-party tip. Vendor limits therefore describe a pipette-plus-tip pair, not a pipette alone, which is the single most common misreading of a datasheet.

The gravimetric reference test

ISO 8655-6 is the gravimetric procedure. Under the 2022 edition the Select Science summary, which draws on the standard's authors, lists these conditions: test room at 20 ± 3 °C, a temperature change of no more than 0.5 °C during the test, water within ±0.5 °C of the room air, relative humidity of 45 to 80 percent, and at least 3 test volumes with at least 10 measurements at each. The three volumes are 100, 50 and 10 percent of nominal, or the minimum volume if that is above 10 percent. Mettler Toledo says Part 6 allows no deviations, and that tips are changed after five dispenses, which means six tips for a 3 × 10 test of a single-channel pipette, or 30 dispenses. The 2002 edition allowed conformity claims on fewer than 10 measurements, and the Mettler quick check lists 4 repetitions in its short procedure. The standard error of a mean with 10 values is 0.63 times that with 4.

Balance requirements: Gilson states that a six-place balance is required for nominal volumes below 20 µL and a five-place balance for 20 to 199 µL. A six-place balance reads 0.001 mg, which is 1 µg. The random limit of a 10 µL pipette is 0.08 µL, about 0.080 mg of water at 20 °C, so the balance reading step is 1.3 percent of it. For a 50 µL pipette the random limit of 0.2 µL is 0.199 mg, and a five-place balance at 0.01 mg is 5.0 percent of it. Select Science notes that the Part 6 environmental conditions are not always achievable on a customer site, which is why Part 7 gives alternative procedures, and Part 8 adds a photometric method.

Z-factor and temperature

The Mettler quick-check table lists the Z factor for distilled water at 1 atm in µL per mg: 1.0020 at 15 °C, 1.0029 at 20 °C, 1.0040 at 25 °C and 1.0054 at 30 °C. A 1000 µL delivery therefore weighs 998.0 mg at 15 °C, 997.1 mg at 20 °C, 996.0 mg at 25 °C and 994.6 mg at 30 °C. The A&D balance manual states that a 5 °C change in water temperature changes the result by 0.11 percent, which matches the 20 to 25 °C step of 0.110 percent computed from the table. That is 1.1 µL on a 1000 µL delivery, or 14 percent of the 8 µL systematic limit, and 0.011 µL on a 10 µL delivery, or 9 percent of the 0.12 µL limit. The same manual says a barometric pressure change from 950 to 1050 hPa changes the result by 0.01 percent.

What each limit means at the bench

A 1000 µL pipette set to 1000 µL may deliver 992.0 to 1008.0 µL on average, an interval of 16.0 µL. Set to 100 µL it may deliver 92.0 to 108.0 µL on average, the same 16.0 µL interval, which is ±8.0 percent. A 100 µL pipette set to 100 µL has an interval of 99.2 to 100.8 µL, 1.6 µL wide.

Random limit in mass at 20 °C, Z of 1.0029 µL per mg: 1000 µL pipette, 3.0 µL is 2.991 mg; 100 µL pipette, 0.3 µL is 0.299 mg; 10 µL pipette, 0.08 µL is 0.080 mg; 2 µL pipette, 0.04 µL is 0.040 mg. Systematic limit in mass at the same Z: 1000 µL pipette, 8.0 µL is 7.977 mg; 100 µL pipette, 0.8 µL is 0.798 mg; 10 µL pipette, 0.12 µL is 0.120 mg; 2 µL pipette, 0.08 µL is 0.080 mg.

Multichannel limits under the 2002 rule are double the single-channel values. For a 1000 µL multichannel the systematic limit is ±1.6 percent, ±16 µL, and the random limit ±0.6 percent, ±6 µL. For a 100 µL multichannel they are ±1.6 µL and ±0.6 µL. For a 10 µL multichannel they are ±0.24 µL and ±0.16 µL, which is 2.4 and 1.6 percent. The 2022 edition replaces this doubling with a separate table.

Water mass at 20 °C for the three test volumes, Z of 1.0029 µL per mg: 1000 µL pipette, 1000 µL is 997.11 mg, 500 µL is 498.55 mg, 100 µL is 99.71 mg. 100 µL pipette: 100 µL is 99.71 mg, 50 µL is 49.86 mg, 10 µL is 9.971 mg. 10 µL pipette: 10 µL is 9.971 mg, 5 µL is 4.986 mg, 1 µL is 0.997 mg.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1601438541475 during review. CPS resolution does not verify the seller's technical claims, the volume range, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

Two things in that listing need settling before it means anything. The model name belongs to a different manufacturer's range while the brand field is empty, so ask who actually makes the pipette and whether the specification sheet is theirs. And because ISO 8655 treats instrument and tip as one system, a claimed limit only holds with the tip it was tested with, so ask which tip was used, which edition of the standard was applied, and at what percentage of nominal the figures are given.

Calibration intervals and users

ISO 8655-1:2022, clause 6.1 in the DIN edition, covers the frequency of calibrations and tests; the text visible did not fix an interval. Sartorius suggests a minimum maintenance interval of one year, and calibration annually or more often, such as every 3 to 6 months, which is between two and four tests a year. Its listed factors are risk of application, frequency of use, number of users, and type of liquid and its vapours, with volatile liquids needing more frequent maintenance. Mettler Toledo says Part 10 covers user qualification, training and competence assessment.

Two points on scope. A test against the ISO limits is what is needed to claim ISO conformity, and a calibration or test can also be run against manufacturer or user tolerances instead. A test against the ISO limits at 10 measurements per volume carries lower uncertainty than one at 4, which Mettler Toledo gives as the reason for the ten-measurement requirement.

What this article could not check

The full text of ISO 8655-2:2022 and its limit tables by volume were not available, so the 2022 numbers appear only as the few rows quoted above. No ISO-stated calibration interval in months was found, and no independent data on how often field pipettes fail the ISO limits. The Mettler Toledo and Select Science summaries come from instrument and service vendors, and the 2002 table was read from a vendor copy and a scanned copy rather than the purchased ISO text.