Laboratory Water Grades: ASTM D1193 Type I to IV, ISO 3696 Grade 1 to 3, CLSI CLRW, and How Fast 18.2 Falls in Air

ASTM Type I asks for 18.0 MΩ·cm, a display shows 18.2, ISO Grade 1 is 10 and CLRW is 10 — and a beaker left open falls to 4 within the hour.

Updated November 21, 2026How Equipvia researches
A laboratory water purification station with mixed-bed and carbon cartridges
Resistivity is measured at the point of use because it falls as soon as the water meets air. Photograph via Wikimedia Commons (CC BY-SA 3.0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

ASTM Type I minimum: 18.0 MΩ·cm. Most system displays: 18.2 MΩ·cm. ISO 3696 Grade 1 maximum: 0.1 µS/cm, which is 10 MΩ·cm. CLSI clinical grade minimum: 10 MΩ·cm. The same word, ultrapure, gets used for all of them. Below, the published limits are set side by side and the reciprocals are checked.

ASTM D1193, four types

The current edition is D1193-24, updated 13 November 2024, replacing D1193-06 (2018); only its abstract could be read here, so the values below come from D1193-06 as tabulated by a water-system maker and by a standards summary, plus older editions.

At 25 °C:

Type Resistivity Conductivity TOC Sodium Silica Chloride
I ≥ 18.0 MΩ·cm ≤ 0.056 µS/cm 50 µg/L 1 µg/L 3 µg/L 1 µg/L
II ≥ 1.0 MΩ·cm ≤ 1.0 µS/cm 50 µg/L 5 µg/L 3 µg/L 5 µg/L
III ≥ 4.0 MΩ·cm ≤ 0.25 µS/cm 200 µg/L 10 µg/L 500 µg/L 10 µg/L
IV ≥ 0.2 MΩ·cm ≤ 5.0 µS/cm no limit 50 µg/L — 50 µg/L

Type IV also carries a pH window of 5.0 to 8.0. The chloride column comes from the maker's table; the standards summary does not list chloride at all.

Preparation, as quoted for the 2018 edition: Type I is distillation or an equal process, then a mixed bed of ion exchange and a 0.2 µm membrane filter; Type II is distillation in a still that gives under 1.0 µS/cm; Type III is distillation, ion exchange, continuous electrodeionization or reverse osmosis, then a 0.45 µm filter. From 2018, other technologies are allowed if the constituent limits are met, including bacteria and endotoxin where they apply.

Reciprocal check

Resistivity in MΩ·cm and conductivity in µS/cm are reciprocals, and the tables agree to the rounding:

  • Type I: 1 / 18.0 = 0.0556 against 0.056 published.
  • Type II: 1 / 1.0 = 1.0 against 1.0.
  • Type III: 1 / 4.0 = 0.25 against 0.25.
  • Type IV: 1 / 0.2 = 5.0 against 5.0.

The conversion table worth memorising: 0.055 µS/cm is 18.2 MΩ·cm, 0.1 is 10, 0.25 is 4.0, 0.5 is 2.0, 1.0 is 1.0, 1.3 is 0.77, 2.1 is 0.48, 4.3 is 0.23 and 5.0 is 0.2.

ISO 3696, three grades

ISO 3696:1987 covers the analysis of inorganic chemicals; its scope excludes organic trace analysis, surface-active agents and biological or medical analysis, and ISO lists the edition as reviewed and confirmed in 2023.

Grade Conductivity Silica Absorbance at 254 nm Residue after evaporation
1 0.1 µS/cm 0.01 mg/L 0.001 —
2 1 µS/cm 0.02 mg/L 0.01 1 mg/kg
3 5 µS/cm — — 2 mg/kg

Grade 3 also has an oxidizable matter limit of 0.4 mg/L. Two entries differ between sources: Grade 3 pH is 5.0 to 7.5 in the standard's own text and in one summary, but 5.0 to 7.0 in two vendor tables; and Grade 2 oxidizable matter as oxygen is 0.008 mg/L in one table against 0.08 in another — a factor of 10. Neither is resolved here, since only a fragment of the standard was read.

In resistivity terms, Grade 1 is 10 MΩ·cm, Grade 2 is 1 MΩ·cm and Grade 3 is 0.2 MΩ·cm.

CLSI, clinical laboratory reagent water

CLSI GP40-A4-AMD defines Clinical Laboratory Reagent Water and Special Reagent Water. A summary of it notes that earlier silicate limits were removed, that CLRW resembles the old Type I, and that CLRW is not required if the laboratory documents reliable results with another grade.

CLRW values by source: resistivity above 10 MΩ·cm with bacteria below 10 CFU/mL; TOC below 500 ppb through a 0.22 µm filter, which the sheet calls a process specification; and a summary giving TOC about 500 µg/L and bacteria about 10 CFU/mL. Special Reagent Water is quoted at TOC below 5 to 10 ppb. Water from reverse osmosis plus electrodeionization typically measures above 10 MΩ·cm and below 50 ppb TOC off-line.

Putting the three side by side

  • ISO Grade 1 conductivity of 0.1 µS/cm is 1.8 times the ASTM Type I figure of 0.056.
  • ISO Grade 1 silica of 10 µg/L is 3.3 times the Type I figure of 3 µg/L; Grade 2 at 20 µg/L is 6.7 times the Type II figure.
  • CLRW's 10 MΩ·cm is exactly ISO Grade 1 at 0.1 µS/cm.
  • ASTM Type I TOC of 50 µg/L is 10 times the 5 ppb a system maker expects from its own product water.
  • Type IV sodium of 50 µg/L is 50 times Type I; Type III is 10 times and Type II 5 times, and chloride follows the same ratios.
  • TOC, Type III against Type I: 4. Silica, Type III against Type I: 167. Conductivity, Type IV against Type I: 89.

The same summary makes the point that ISO Grade 1 does not automatically meet ASTM Type I, and that "ultrapure" is not a formal grade.

Bacteria and endotoxin

Older ASTM editions attach bacteria and endotoxin limits to microbiological classes A, B and C, and two sources disagree on class C.

Class D1193-91 D1193-99e1
A 10 per 1,000 mL 10 per 1,000 mL
B 10 per 100 mL 10 per 100 mL
C 10 per 100 mL 100 per 10 mL

Endotoxin is quoted under 0.03 EU/mL for class A and 0.25 for class B, which at roughly 0.1 ng per EU is under 0.003 and 0.025 ng/mL. In per-litre terms: class A is 10 per L, class B 100 per L, class C 10,000 per L, CLRW at 10 CFU/mL is 10,000 per L, and USP purified water at 100 CFU/mL is 100,000 per L. Class C differs by a factor of 1,000 between the two editions, so the edition matters if a method cites it. No numeric bacteria table was found for the 2024 edition.

What happens in a beaker

Pure water at 25 °C is 18.18 MΩ·cm in theory, and 0.055 µS/cm corresponds to 18.2 with no CO₂ present. Once it meets air, the readings diverge by source:

Source Condition Reading
System maker A CO₂ up to 500 ppm 1 µS/cm, about 1 MΩ·cm
System maker B about one hour 18.2 down to 4 MΩ·cm
System maker C minutes to one hour 10 to 12 MΩ·cm
System maker C a few hours below 10 MΩ·cm

The three do not agree: at about an hour, maker B's figure and maker C's differ by a factor of 2.5 to 3, and neither publishes the measurement conditions.

Converted: 4 MΩ·cm is 0.25 µS/cm, exactly the Type III conductivity ceiling; 10 to 12 MΩ·cm is 0.100 to 0.083 µS/cm, inside ISO Grade 1 and CLRW but outside Type I; 1 MΩ·cm is the Type II minimum. Maker B's drop is 4.6 times in an hour, maker C's 1.5 to 1.8 times. Maker A's 1 µS/cm is 18 times the 0.055 baseline.

The handling advice that follows: do not store ultrapure water, expect TOC at or below 5 ppb from the system, and treat all specifications as applying at the point of use.

Where 18.0 and 18.2 sit

Pure water at 25 °C is 18.18 MΩ·cm, or 0.0550 µS/cm. A display of 18.2 is 0.0549 µS/cm, and the ASTM Type I minimum of 18.0 is 0.0556 µS/cm. The headroom is (18.18 − 18.0) / 18.18 = 1.0 percent, or 0.0006 µS/cm — the whole difference between a "perfect" 18.2 and a passing Type I is smaller than the meter's own drift.

What the numbers mean for a purchase

Grade Resistivity Conductivity TOC Bacteria
ASTM Type I 18.0 MΩ·cm 0.056 µS/cm 50 µg/L class A
ISO Grade 1 10 MΩ·cm approx. 0.1 µS/cm not listed not listed
CLRW above 10 MΩ·cm 0.1 µS/cm 500 ppb below 10 CFU/mL
USP purified water 0.77 MΩ·cm 1.3 µS/cm 500 ppb 100 CFU/mL

Resistivity does not detect TOC, bacteria, endotoxin or particles. A single-pass distillate is not a defined ASTM or ISO grade. A system labelled Type I should name the standard and the edition, because the same label covers 0.056 and 0.1 µS/cm.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 10000041637891 during review. CPS resolution does not verify the seller's technical claims, the TOC or bacteria figures, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

18.2 MΩ·cm is the number every ultrapure system displays, and it is also the number that proves the least: it says nothing about TOC, bacteria, endotoxin or particles, and it collapses within an hour in an open beaker. For HPLC work the TOC figure and the 0.22 µm filter matter more; for ICP-MS it is the trace-metal and silica limits; for cell work it is endotoxin. Ask which standard and edition the claim refers to, and ask for TOC at the point of use rather than at the outlet.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1601779517911 during review. CPS resolution does not verify the seller's technical claims, the grade, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

A 30 L/h RO/DI unit sits upstream of the grade question: it produces feed water and rinse water, and whether it reaches Type II depends on the resistivity at the outlet, which the listing does not give. Type II allows 1.0 µS/cm, or 1 MΩ·cm — a figure a two-stage RO plus DI can reach and a single-stage unit often does not. Ask for the outlet resistivity, the TOC if any, and which cartridges are included, because the mixed bed is the part that sets the number.

Pharmacopoeial water, for comparison

USP conductivity: Stage 1 at 1.3 µS/cm at 25 °C for purified water and water for injection, unequilibrated; Stage 2 at 2.1 µS/cm, equilibrated. TOC under <643> is 500 ppb, and the suggested microbial level under <1231> is 100 CFU/mL for purified water. The USP 35 text of <645> specifies a Stage 2 sample at 25 ± 1 °C, agitated until the change is under 0.1 µS/cm per 5 minutes.

Ph. Eur. gives purified water 4.3 µS/cm at 20 °C and water for injection 1.1 µS/cm at 20 °C, with bacterial endotoxin for WFI at 0.25 IU/mL and aerobic bacteria at 10 CFU per 100 mL.

Conversions: USP Stage 1 is 0.77 MΩ·cm and Stage 2 is 0.48; Ph. Eur. purified water is 0.23 MΩ·cm, though at 20 °C rather than 25 °C, so it does not compare directly. USP Stage 1 is 23 times the ASTM Type I conductivity, 1.3 times Type II, 5.2 times Type III and 13 times both ISO Grade 1 and CLRW; its TOC is 10 times the Type I limit. Ph. Eur. purified water against WFI is 3.9 times, and USP Stage 2 against Stage 1 is 1.6.

Containers and handling

Discard about 1 to 2 litres before collecting for sensitive work, since a system idle overnight can pick up airborne volatiles. Use borosilicate glass for organic analyses such as HPLC, polyethylene or polypropylene for ionic analyses, and PFA for ICP-MS. Avoid plastic tubing at the outlet, because plasticizers can leach. Do not store the water: dispense it just before use. The same note gives no storage time limit and no rinse count.

What this article could not check

The text of ASTM D1193-24 beyond its abstract was not available, nor ISO 3696 beyond a fragment — so the Grade 3 pH and Grade 2 oxidizable matter conflicts stand — nor the text of CLSI GP40-A4-AMD beyond summaries, nor a numeric bacteria and endotoxin table for the 2024 ASTM edition, nor the test conditions behind the two air-exposure figures.