Cryogenic Vials: Internal or External Thread, and the 694-to-1 Expansion
A sealed vial holding 0.1 mL of liquid nitrogen reaches about 58 atm inside a 1.2 mL tube at room temperature. That is the arithmetic nobody puts on the box.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.
Thermo Fisher publishes two answers about thread type for its Nunc cryotubes. One FAQ says internally threaded tubes with a silicone gasket are designed for vapor-phase liquid nitrogen and externally threaded tubes for mechanical freezers. Another FAQ for the same range says both types are designed and tested for cryogenic storage and the choice is largely preference. Yale Environmental Health and Safety describes an external-thread vial with a female cap that exploded in April 2012 and a nearly identical event in July 2020. Fisher Scientific's own vial pages say the Fisherbrand vials are intended for the vapor phase only. Air Products gives liquid nitrogen an expansion ratio of 1 to 694 from boiling point to 20 °C, so a sealed vial that holds even a small amount of liquid becomes a pressure vessel on warming.
What the vendors and safety offices say about thread type
Thermo Fisher's internal-thread FAQ recommends the internal thread with a gasket for liquid nitrogen immersion and adds that any mechanical closure carries a theoretical leak risk. Its external-thread FAQ gives a different reason for that design: the external cap reduces contamination from handling. Fisher Scientific's Italian FAQ says the same, that external threads give a secure seal and lower the contamination risk. A distributor page from Genfollower says internal-thread vials are generally preferred when the aim is to minimise liquid nitrogen penetration and external-thread vials suit fast cap handling and automated capping.
Thermo Fisher's gasket application note describes a leak test at 5 inHg vacuum for 30 minutes after 10 freeze and thaw cycles in liquid nitrogen; 5 inHg is 16.9 kPa. One aged 1.8 mL Nunc tube failed, and re-tightening fixed it. Yale advises vapor phase only and internal threads with male caps where possible. UAB says male caps on internal threads are safer than female caps on external threads.
Counting positions: three sources for internal thread (Yale, UAB, Genfollower), two that tie external thread to lower handling contamination (Thermo's external-thread FAQ, Fisher Italy) and one Thermo FAQ that calls the choice preference. The Institut für Luft- und Kältetechnik Dresden paper at the Cryogenics 2021 conference does not compare internal and external threads at all. Its gravimetric test used external-thread tubes from one manufacturer, and that is the only thread detail in it.
Measured liquid nitrogen leakage
BioLife Solutions, a vendor, lists four percentages in its white paper. Two come from Clarke 1999 as cited there, in a 3-hour immersion: 45 percent for Nunc 340711 without an O-ring and 85 percent for Iwaki with an O-ring. Two are BioLife's own: 92 percent for Corning 430488 and 18 percent for 340711. The ratio 85 / 45 is 1.89 and the ratio 92 / 18 is 5.11. Sample sizes are not given in the text read, and catalogue 340711 is Nunc in one passage and Nalgene in another. These figures could not be traced to a peer-reviewed table.
The Dresden study used three tests. The carbon dioxide test simulated dry-ice transport with 96 vials from each of two manufacturers, and one manufacturer's tubes met the paper's tightness criteria while the other's leaked. The gravimetric test followed an IATA-based protocol from −40 to +55 °C at about 99 kPa test pressure on 96 fresh and 96 thermally stressed tubes, and some previously tight tubes leaked after stress. For liquid nitrogen the test showed complete tightness for 0 of the tubes tested, and the text gives no leak counts.
Tensile data in the same paper: for some products, maximum tensile strength fell from 410.3 ± 3.3 N to 392.3 ± 5.4 N after 50 thermal cycles, a drop of 18.0 N or 4.4 percent. Other manufacturers showed no significant change. For straws stored since 1995 the paper gives 14.5 ± 0.3 N against 16.4 ± 0.7 N fresh; the table gives 11.6 percent lower while the text gives almost 15 percent lower, and the 2006 straws at 14.3 N are 12.8 percent lower.
Chen and colleagues (Cryobiology, 2006) tested a heat-sealable membrane over the vial top and report that it prevented liquid nitrogen penetration without changing cell viability. Thermo Fisher's route to the same end is the CryoFlex tube wrap, catalog 343958, a heat-sealed barrier that it says is required if tubes go into the liquid phase. Thermo also explains why a tight cap alone does not settle the matter: liquid nitrogen has very low surface tension and can flow into a vial through a closed cap.
The 694-to-1 arithmetic for 1.2, 2 and 5 mL vials
Air Products Safetygram 7 lists liquid nitrogen at a boiling point of −195.8 °C, a liquid density of 808.5 kg/m³ and a latent heat of 199.1 kJ/kg, with an expansion ratio of 1 to 694 from boiling point to 20 °C. That density is 0.8085 g/mL, so 0.1 mL of liquid is 80.9 mg and takes 16.1 J to vaporise. The sources do not agree on the ratio: USDA, Creighton and Wikipedia are quoted as 694, Utah State University EHS gives 696, the National University of Singapore gives 682.1 at 15 °C and St George's, University of London gives 683. The spread from 682.1 to 696 is 13.9, which is 2.0 percent of 682.1, and a change of that size does not alter any pressure below by more than 2 percent.
The table assumes a sealed vial, trapped liquid nitrogen, warming to 20 °C, no leak, ideal gas and the ratio 694. Burst pressure of polypropylene vials or caps was not found in the sources read, so the table shows pressure only.
| Trapped liquid | Vial volume | Fill fraction | Nitrogen gas at 20 °C | Gas pressure |
|---|---|---|---|---|
| 0.05 mL (40.4 mg) | 1.2 mL | 4.2 % | 34.7 mL | 28.9 atm (2.93 MPa) |
| 0.1 mL (80.9 mg) | 1.2 mL | 8.3 % | 69.4 mL | 57.9 atm (5.86 MPa) |
| 0.2 mL (161.7 mg) | 1.2 mL | 16.7 % | 138.8 mL | 115.7 atm (11.72 MPa) |
| 0.05 mL | 2 mL | 2.5 % | 34.7 mL | 17.4 atm (1.76 MPa) |
| 0.1 mL | 2 mL | 5.0 % | 69.4 mL | 34.7 atm (3.52 MPa) |
| 0.2 mL | 2 mL | 10.0 % | 138.8 mL | 69.4 atm (7.03 MPa) |
| 0.05 mL | 5 mL | 1.0 % | 34.7 mL | 6.9 atm (0.70 MPa) |
| 0.1 mL | 5 mL | 2.0 % | 69.4 mL | 13.9 atm (1.41 MPa) |
| 0.2 mL | 5 mL | 4.0 % | 138.8 mL | 27.8 atm (2.81 MPa) |
The arithmetic behind each row is gas volume divided by vial volume: 34.7 / 1.2 is 28.9, 34.7 / 2 is 17.4 and 34.7 / 5 is 6.9. The same 0.05 mL of liquid gives 28.9 atm in a 1.2 mL vial and 6.9 atm in a 5 mL vial, and 28.9 / 6.9 is 4.19 with rounding. Latent heat at 199.1 J/g is 8.0 J for 0.05 mL, 16.1 J for 0.1 mL and 32.2 J for 0.2 mL, and a 1.2 mL vial filled completely holds 0.97 g of liquid, which is 193 J. A carrier at −150 °C is 123.15 K and −238 °F, 15 K below the ISBER figure of −135 °C.
USU EHS attributes exploding vials to the 1:696 expansion. Yale's April 2012 case involved an external-thread vial with a female cap. Yale advises cryoflex tubing for immersion, and for vials that have been immersed it advises placing them in a sealed unbreakable container before thawing. The Fisher vial pages add: never overfill the storage unit, inspect closure rims, wear a face shield and heavy gloves, warm slowly in a biological safety cabinet, and never reuse a vial.
Vapor phase temperatures and glass transition
Published vapor-phase temperatures differ by source. SciMed and PHC quote −130 °C at the warm end. Other pages quote −135 to −190 °C, and isothermal liquid nitrogen freezers are described as holding about −190 °C. A CryoPod carrier is described as holding −150 °C. Liquid nitrogen itself at −196 °C is 77.15 K and −320.8 °F.
The temperature below which stored biological material is treated as inert is also a range. ISBER, via Azenta, gives about −135 °C, the Thermo Fisher blog gives −132 °C, and a Chromatography Today article describing an Azenta study gives −134 °C. In kelvin these are 138.15, 141.15 and 139.15, a spread of 3 K. A −130 °C vapor-phase top end (143.15 K) is 5 K warmer than the ISBER figure and 2 K warmer than the Thermo figure, so on the published numbers the warm end of the vapor range sits above all three thresholds. The −80 °C mechanical freezer (193.15 K, −112 °F) is 55 K above the ISBER figure. A Tissue Engineering Part C study from 2013 found fewer viable cells after one month at −80 °C than at −170 °C. Azenta states that dry ice is about 90 °C warmer than the liquid nitrogen vapor environment. In Fahrenheit the thresholds are −211 °F (−135 °C), −209.2 °F (−134 °C) and −205.6 °F (−132 °C), and the −130 °C vapor top end is −202 °F.
Azenta's thermal excursion study reports that handling below −150 °C cuts absorbed heat by more than 90 percent and that the interior temperature spread stays within 4 °C. Its thermal diffusivity values are 1.07 × 10⁻⁵ mm²/s for ice at −150 °C and 1.29 × 10⁻⁷ mm²/s for polypropylene at 22 °C. The ratio is 82.9, but the two numbers are at different temperatures and in different materials.
Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1600764271685 during review. CPS resolution does not verify the seller's technical claims, the storage rating, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.
Two words in that listing need separating. It is titled as both a cryovial and a self-standing transport tube, and those are different products with different testing: a transport tube needs a leak-proof closure for carriage, while a cryovial needs a closure tested after freeze-thaw cycles in liquid nitrogen. The internal thread is the safer pattern for immersion, and the 10 mL size is more typical of a transport tube than of a cryovial. Ask which phase the tube is rated for, whether the closure was leak-tested in liquid nitrogen and by what method, and whether the sterility claim covers nucleases, since a tube chosen for the wrong phase is exactly the case the pressure table above describes.
Removal and thawing steps in the published guidance
Thermo Fisher's protocol for tubes that were immersed is at least 24 hours in the gas phase, then a water bath under a towel. UAB says to move vials from liquid to vapor phase 24 to 48 hours before removal. The two ranges overlap at 24 to 48 hours. UAB adds that polypropylene is the best plastic for liquid nitrogen, that glass and polystyrene crack, and that caps should not be overtightened because the gasket is damaged. Thermo says its tubes should not be autoclaved.
Three controls appear across those documents: closure style, storage phase and thaw procedure. Fisher's vial pages exclude the liquid phase for their own products. Yale and Thermo Fisher allow immersion only with a heat-sealed wrap. UAB and Yale both ask for a 24 to 48 hour vapor hold before removal.
Contamination records
Tedder and colleagues (The Lancet, 1995; 346:137 to 140) report that 6 patients acquired hepatitis B from a contaminated cryopreservation tank, where bag leakage had contaminated the liquid nitrogen. Hawkins and colleagues (Journal of Virological Methods, 1996) used sequence analysis to link the cluster. Bielanski and colleagues (Cryobiology, 2000; 40(2):110 to 116) found that sealed cryovials and straws stayed free of virus in contaminated liquid nitrogen. Bielanski and colleagues (Cryobiology, 2003; 46(2)) found 32 bacterial species and 1 fungal species in 8 commercial and 8 research tanks, with storage periods of 6 to 35 years, across 16 tanks.
Grout and Morris 2009 (Theriogenology 71:1079 to 1082) deals with contaminated liquid nitrogen vapor. Schenkel and Overmann 2019 (Applied Microbiology and Biotechnology) covers tank colonisation and advises hermetically sealed containers. The Human Reproduction Open table hoab037 lists vapor-phase storage and wiping vials with a disinfectant after liquid nitrogen removal. Three sources, three different recommendations.
What the article leaves open
Published: the 694 to 1 ratio, the 24 to 48 hour hold, and the −132 to −135 °C thresholds. Arithmetic: every pressure in the table, from PV = nRT at 293.15 K with n equal to mass divided by 28.0134 g/mol. Not found in any source read: the burst pressure of a named vial, leak counts by thread type, BioLife's sample sizes, and a thread comparison at one material and one volume. ISBER Best Practices and its 2019 addendum were read as summaries only.
What this article could not check
The ISBER text was not available in full, so the threshold figures come from secondary citations. The BioLife percentages could not be traced to a peer-reviewed table and their sample sizes are unknown. The Dresden paper's liquid nitrogen test reports that no tube was completely tight, without counts, so it cannot be used to rank products. No burst-pressure data for a specific vial or cap was found, which is why the pressure table stops at pressure and does not say at what point a tube fails.