Selection GuideRotary evaporators15 min read

Rotary Vane Pump Oil: Mineral, Di-ester or PFPE, and What the Vapour Pressure Numbers Actually Say

The mineral oils in one maker''s range differ by a factor of 27.6 in vapour pressure. Picking the cheapest one is a decision about ultimate vacuum, not about price.

Updated November 17, 2026How Equipvia researches
A cutaway laboratory oil rotary vane vacuum pump showing its internal rotor and oil reservoir
The oil seals, lubricates and carries away heat in the same chamber, which is why its vapour pressure sets the pump''s floor. Photograph via Wikimedia Commons (CC0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

Edwards gives its mineral rotary vane oils vapour pressures between 2.1×10⁻⁷ and 5.8×10⁻⁶ mbar at 20 °C, a spread of 27.6 times. Its di-ester oil is 5×10⁻⁵ mbar at 20 °C. Its PFPE oils run from 1.2×10⁻⁷ to 1.07×10⁻⁶ mbar at 25 °C. Below, those figures are put next to VACUUBRAND, Welch, Halocarbon and Pfeiffer, with Pa and torr conversions, and the oil-change signs come from Leybold.

What the oil has to do

Leybold lists lubricating the moving parts among the jobs of the oil in a rotary vane pump. A Fabrico troubleshooting guide for oil-sealed and dry vane pumps puts the sealing role in one line: vacuum pump oil is a sealing fluid, not just a lubricant. Vacaero adds the vapour side. If the oil vapour pressure is too high, the oil evaporates under vacuum, and the vapour can backstream into the chamber. Vapour pressure is one of the factors that sets the ultimate vacuum.

Two requirements recur in the sources: a sufficiently low vapour pressure and the correct viscosity. Vacaero's rule of thumb for sizing is that a pump should reach at least one decade below the required process vacuum. For a process at 1 mbar that is 0.1 mbar; for a process at 10⁻² mbar it is 10⁻³ mbar.

Mineral oils, Edwards Ultragrade and Xtragrade

The Edwards oil overview flyer gives vapour pressure at 20 °C and viscosity at 40 °C. Conversions to pascal and torr are arithmetic, and the last column is the ratio to Performance 20.

Oil (Edwards) Vapour pressure, mbar Pa Viscosity, cSt at 40 °C Ratio to Performance 20
Performance 15 5.8×10⁻⁶ 5.8×10⁻⁴ 38.1 27.6
Performance 19 1.0×10⁻⁶ 1.0×10⁻⁴ 48.6 4.8
Performance 20 2.1×10⁻⁷ 2.1×10⁻⁵ 103 1.0
Performance 70 2.1×10⁻⁶ 2.1×10⁻⁴ 70 10.0
V-Lube B 5.0×10⁻⁶ 5.0×10⁻⁴ 94.3 23.8
V-Lube F 1.3×10⁻³ 0.13 105.4 6,190

Edwards assigns the grades by pump class. Performance 15 and 19 are for small rotary vane pumps of 0.5 to 5 m³/h. Performance 20 is listed for gearboxes and drive trains under vacuum and for large single-stage rotary vane pumps. Performance 70 is for large two-stage pumps under high load or fast duty cycles. V-Lube B is for single-stage rotary vane pumps that need a heavier oil for the vacuum seal. V-Lube F is for oil piston pumps, a different pump class, and its vapour pressure is 6,190 times that of Performance 20, which is the clearest illustration in the table that these grades are not interchangeable.

Welch lists Labovac 10 mineral oil for its two-stage rotary vane pumps and Chemvac pumps: 10⁻⁶ mbar, 118 cSt at 40 °C, a flash point of 270 °C and a density of 0.888 g/mL at 15 °C, for air, inert gases and noble gases. At 118 cSt it is 3.1 times the Performance 15 viscosity and 1.15 times Performance 20.

Di-ester and synthetic oils

Edwards Endurance EXTEND 110 is a di-ester oil: 5×10⁻⁵ mbar at 20 °C, 98 cSt at 40 °C. Its vapour pressure is 238 times that of Performance 20, which reads badly until the use case is added: Edwards lists it for large and small rotary vane pumps in high-temperature service, where a mineral oil would oxidise faster than it would evaporate. Welch's Labovac 11 is a synthetic with 10⁻⁵ mbar for operating temperatures above 100 °C, especially in one-stage rotary vane pumps, which is 47.6 times the Performance 20 figure.

Edwards TW 45 is a special synthetic at 1.33×10⁻⁶ mbar and 25 °C, listed at 65 and 56 cSt. Edwards calls it heavily distilled with no additives, for reactive or corrosive substances. Kinetic 46 is a PAO synthetic at 1.39×10⁻⁶ mbar and 20 °C, 44 cSt, for oil-sealed screw pumps; its vapour pressure differs from TW 45 by 4.5 percent, but the two are quoted at different temperatures.

PFPE oils

Edwards lists four PFPE grades at 25 °C, with viscosity given at 20 °C rather than 40 °C.

  • PFPE 64: 1.07×10⁻⁶ mbar, 64 cSt. Small and medium rotary vane pumps.
  • PFPE 148: 1.3×10⁻⁷ mbar, 148 cSt. Mechanical boosters.
  • PFPE 168: 1.2×10⁻⁷ mbar, 168 cSt. Large rotary vane pumps and boosters.
  • PFPE 276: 1.33×10⁻⁷ mbar, 305 cSt. Dry screw pumps.

All four are listed for the same reactive gases: oxygen, ozone, halogens, hydrogen, UF₆, F₂ and tritium. PFPE 64 sits at 8.0×10⁻⁷ torr, 5.1 times the Performance 20 figure; PFPE 168 sits at 9.0×10⁻⁸ torr, 0.57 times it. So the PFPE family is not uniformly better on vapour pressure — the low-viscosity grade is worse than the best mineral oil, and the reason to buy it is chemical inertness, not vacuum.

VACUUBRAND Oil B has a vapour pressure below 1×10⁻³ mbar at pump operating temperature, a flash point of 264 °C, a viscosity of 94 mm²/s at 40 °C and a density of 0.87 g/cm³. Oil K 8 is below 5×10⁻³ mbar, 249 °C, 128 mm²/s and 0.89 g/cm³. PFPE type I and type II are below 3×10⁻⁵ mbar, with 60 mm²/s and 1.90 and 1.89 g/cm³. The PFPE limit is 33 times lower than Oil B's, and PFPE is 2.18 times denser, which matters when a fill volume is specified by mass.

Three published rules for choosing by gas

Pfeiffer states that mineral oils work well for air, noble gases and water vapour, and that PFPE oils, being chemically inert, are for corrosive processes. It adds that mineral oil, or a mineral and synthetic mixture, is enough for most applications. VACUUBRAND describes Oil B as having good chemical resistance, with better stability than standard mineral oil against acid and alkaline vapours. It describes K 8 as designed for acid vapours but very hygroscopic, with limited water vapour capacity and a consumed alkaline additive, so it must be changed regularly; pumps may not reach the specified ultimate vacuum on K 8 and may start badly below 18 °C.

VACUUBRAND states that PFPE must not be mixed with mineral oils. A pump must be built for it, or be completely disassembled, cleaned and refilled to convert. PFPE type I is intended for strong oxidants such as halogenides and nitrogen oxides, for the RE 2.5, RZ 2.5, RE 6 and RZ 6 models.

Leybold adds two limits. Its corrosion protection oil, PROTELEN, is for corrosive vapours such as acids, but is not recommended for normal operation because it absorbs moisture and cannot handle water vapour. Pure oxygen needs specially formulated oil with BAM approval.

Unit conversions for the main grades

Using 1 mbar = 100 Pa and 1 mbar = 0.750 torr: Performance 20 at 2.1×10⁻⁷ mbar is 2.1×10⁻⁵ Pa and 1.58×10⁻⁷ torr. Performance 19 at 1.0×10⁻⁶ mbar is 1.0×10⁻⁴ Pa and 7.5×10⁻⁷ torr. Performance 15 at 5.8×10⁻⁶ mbar is 5.8×10⁻⁴ Pa and 4.35×10⁻⁶ torr. EXTEND 110 at 5×10⁻⁵ mbar is 5×10⁻³ Pa and 3.75×10⁻⁵ torr. PFPE 168 at 1.2×10⁻⁷ mbar is 1.2×10⁻⁵ Pa and 9.0×10⁻⁸ torr. PFPE 64 at 1.07×10⁻⁶ mbar is 1.07×10⁻⁴ Pa and 8.0×10⁻⁷ torr. TW 45 at 1.33×10⁻⁶ mbar is 1.33×10⁻⁴ Pa and 1.0×10⁻⁶ torr. V-Lube F at 1.3×10⁻³ mbar is 0.13 Pa and 9.75×10⁻⁴ torr. Viscosity ratios at 40 °C: Labovac 10 at 118 cSt over Performance 15 at 38.1 cSt is 3.1; Performance 20 at 103 cSt over Performance 15 is 2.7; K 8 at 128 mm²/s over Oil B at 94 mm²/s is 1.4.

Where the temperature bases differ

The sources do not compare like with like. Edwards quotes 20 °C for most oils and 25 °C for TW 45, PFPE and the diffusion fluids. VACUUBRAND quotes the pump operating temperature. Halocarbon quotes mmHg at 40 °C for its HaloVac fluids: 6.3×10⁻² mmHg for HaloVac 60, 2.5×10⁻² for HaloVac 100, 1.9×10⁻² for HaloVac 125 and 7.8×10⁻³ for HaloVac 190, which is 0.084, 0.033, 0.025 and 0.0104 mbar. Those four values are 10⁴ to 10⁵ times the Edwards PFPE figures, but they are measured at 40 °C against 25 °C and describe a different fluid family, so they are not ranked against the Edwards table here.

Halocarbon's page also gives one fluid at several temperatures: 0.0016 mmHg at 25 °C, 0.063 mmHg at 75 °C and 2.9 mmHg at 150 °C. From 25 to 75 °C that is a factor of 39; from 25 to 150 °C it is 1,813. One conflict sits inside the same maker's material: the 0.063 mmHg figure appears for HaloVac 60 at 40 °C on one page and for the first fluid at 75 °C on another, and it could not be established whether these are the same fluid. Vacaero states the general point: vapour pressure rises as the fluid gets hotter, and the pump does not reach quite as low a pressure.

Viscosity bases differ too. VACUUBRAND PFPE is 60 mm²/s at 40 °C; Edwards PFPE 64 is 64 cSt at 20 °C. One mm²/s is one cSt, so the numbers are comparable in unit, not in temperature — and a PFPE that reads thinner at 20 °C may be thicker than a mineral oil at 40 °C.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1601456190672 during review. CPS resolution does not verify the seller's technical claims, the grade, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

The listing does give two useful facts: it is a mineral oil at ISO VG 46 with a viscosity index of 110, which puts it in the same band as Welch's Labovac 10 at 118 cSt rather than among the thin 15-grade oils. What it does not give is the number that decides the ultimate vacuum — vapour pressure at a stated temperature — nor the additive package, which is what separates a pump oil from a hydraulic oil of the same viscosity. Ask for the vapour pressure figure and whether the maker approves the oil for your pump model, because a mineral oil in a pump built for PFPE, or the reverse, damages seals rather than merely costing vacuum. A 330 mL bottle also only covers a top-up on most rotary vane pumps, whose fills run to a litre or more.

Signs that the oil needs changing

Leybold states that there is no fixed interval. Under clean conditions, such as backing pumps for diffusion pumps in accelerators, a rotary vane pump can run for years without an oil change; under extremely dirty conditions, such as impregnation, the oil may need changing daily.

The signs Leybold lists: colour light brown at the start, dark brown or black when aged; cloudiness means water or another liquid is in the pump; flakes in corrosion protection oil mean the inhibitor is used up; process vapours in the oil may make it unusable. Its procedure is pump off, oil drained at operating temperature through the drain opening, level never below the sight-glass minimum in continuous operation. Oil is stiff below 10 °C.

Fabrico gives two matching rows. Milky or foamy oil means water or condensate, so run the pump warm with the gas ballast open and change the oil if it stays emulsified. Dark or burned-smelling oil means overheating or an overdue change, so find the heat source before refilling. Leybold adds that once the oil has absorbed water the pump can no longer reach its ultimate pressure, which is why the colour is a symptom and the ultimate pressure is the measurement.

Condensation protection and oil loss

Leybold lists three options for condensation protection. Gas ballast adds air to the compression chamber and is not for hydrogen. A higher pump temperature, up to 120 °C, gives a water vapour tolerance about 5 times that of normal gas ballast operation. A vacuum condenser upstream removes vapour before the pump.

Without separation and return, Leybold gives oil loss as about 2 cm³ per m³ pumped, worst case, and gas ballast increases it. A coarse oil separator in the discharge line reduces the loss considerably; a fine oil mist filter brings it to practically zero because the separated oil returns to the pump. As an example, a 5 m³/h pump running 8 hours pumps 40 m³, which at 2 cm³ per m³ is 80 cm³ in the worst case, or 8 percent of a 1 L fill. Leybold does not give the pressure at which the 2 cm³ figure applies, so that example is an upper bound.

Blank-off test before blaming the oil

The Fabrico blank-off test has three steps: valve off the pump at its inlet; fit a blank-off flange with a trustworthy gauge on the inlet port; run to operating temperature and record the best vacuum. A result near the rated ultimate pressure in the manual points to a system problem. A result well short points to the pump, with three listed causes: worn vanes, degraded oil and a leaking exhaust valve. Leybold adds that an ultimate pressure degraded by contaminated oil can also indicate worn or damaged internals that need repair.

The ultimate or base pressure itself is a defined measurement. ISO 21360-1:2020 specifies one method for measuring base pressure, three for volume flow rate, and one each for compression ratio and critical backing pressure. The sources read here do not give a base pressure test in the oil-change context beyond the blank-off test above.

What this article could not check

No independent test of ultimate pressure against oil grade on one pump was found, nor any standard fixing one temperature for quoting oil vapour pressure, nor an oil-change interval in hours, nor a base pressure for one pump model on two oils. All figures come from maker and distributor sheets. The Edwards flyer gave two TW 45 value columns under one name and unclear diffusion-fluid labels, so the diffusion fluids are left out.

Share this article

View all guides
Selection Guides·14 min read

What Exhaust Filter Fits a Busch R5 RA 0063 F?

Two Busch part numbers look plausible for the same nameplate. They describe different exhaust-filter arrangements. This guide shows which manual governs, how many elements a standard pump takes, and what to record before ordering.

Updated October 21, 2026